The medical molecular sieve oxygen generator utilizes zeolite molecular sieve as the adsorbent and employs the Pressure Swing Adsorption (PSA) process to produce medical oxygen (also referred to as an oxygen generator).
- The data listed in this table is based on 0.8Mpa raw material compressed air, 20℃ ambient temperature,100 meters above sea level, and 80% relative humidity.
- Power requirement: 380V 3P,220V 1P,50HZ
- Oxygen output pressure requirements: 0.4-0.5Mpa, when the oxygen pressure requirements are higher than 0.5Mpa, can be pressurized to the user's required pressure.
- Installation site requirements: The installation site is an indoor non-explosion-proof area with an ambient temperature of 3-45℃ and good ventilation.
Parameters | International standard ISO 10083 |
USP XXII 93% oxygen |
EP 93% oxygen |
Looming Standard | ||||||||||
O₂ | >90% | 90%-96% | 90%-96% | ≥90% | ||||||||||
CO | <5 ppm | <0.001% | <5 ppm | ≤5×10-6 | ||||||||||
CO₂ | <300 ppm | <0.03% | <300 ppm | ≤0.01% | ||||||||||
CO₂ | - | - | <1 ppm | In accordance with YY/T 0298-1998 | ||||||||||
NOx | - | - | <2 ppm | In accordance with YY/T0298-1998 | ||||||||||
H2O | <67 ppm (-50℃) | - | <67 ppm (-50℃) | ≤0.07g/m³ | ||||||||||
Oil | <0.1mg/m³ | <0.1mg/m³ | <0.1mg/m³ | ≤0.1mg/m³ |